Adimea technology

Progressive UV measurement

The pioneering technology enables measurement in the spent dialysate. Because it continuously analyzes change in the molar concentration, Adimea is directly connected to the patient. On the basis of this characteristic, Adimea is a highly reliable and accurate instrument for an online determination of the current Kt/V value during treatment.

During therapy, the system - at the request of the user - generates a warning message in the event that the planned target value is not reached. This enables the user to carry out target-oriented adjustments to the treatment parameters – at any time during the ongoing dialysis.

Adimea utilizes the principles of spectroscopy for determining the reduction in the molar concentration of urinary excreted substances in the dialysate drain. A light source (1) transmits ultraviolet light (2) through the dialysate. The particles contained in the dialysate, which were removed from the plasma during dialysis, absorb the light. This absorption is measured by a sensor (3) .

 

Principle

It has been proven [1], that the UV absorption measurements can be used for determining the dialysis dose as there is a very close linear correlation between the measured UV absorption signal and the urea in the dialysate. For this reason, the measurement values recorded by Adimea can be used as replacement parameters for the urea.

 

Functionality during the treatment

The progress of the measured substance reduction provides a true overview of the administered dialysis dose. Depending on the effectiveness of the dialysis, the concentration of urinary
excreted substances, and hence also the UV light absorption, reduces over the course of the treatment. The continual measurements result in a patient-specific curve which almost precisely corresponds to the reduction in urea.

Adimea processes this acquired information to determine, and clearly display, the Kt/V during treatment. (see treatment progress graph below)

The dialysis progress becomes apparent quickly and easily. The graphic display of the treatment progress and relevant parameters (selected treatment index, Kt/V and URR, actual values achieved, overview of and access to: treatment time / blood flow / dialysate flow) enables optimum user support at all times.

 

 

_________________________________

[1] Uhlin F, Fridolin I, Magnusson M, Linberg L-G. Dialysis dose (Kt/V) and clearance variation sensitivity using measurement of ultraviolet absorbance (on-line), blood urea, dialysate urea and ionic dialysance. Nephrol Dial Transplant (2006) 21: 2225-2231