ECCOS® Mechanical Reprocessing

Product Quick Finder

Choose a category or subcategory

The solution for storage and mechanical reprocessing of your Aesculap power systems

Basic requirements for mechanical reprocessing in combination with the storage system:

  • A suitable cleaner must be used (pH <11 for aluminum components)
  • The temperature in the cleaning cycle must not exceed 60°C
  • Clean for at least 3 minutes
  • Apply intermediate rinse for at least 1 minute
  • Carry out intensive final rinse with distilled, demineralized or fully desalinated water
  • Only thermal disinfection: 5 minutes rinse at 96°C with distilled, demineralized or fully desalinated water
  • Complete the program with a drying phase of at least 10 minutes at a temperature of not more than 120°