B. Braun ensures high security standards throughout the product life cycle by using globally accepted standards. We have established processes to monitor the latest vulnerabilities, threats, or risks and will proactively implement measures as required.
Successful exploitation of these vulnerabilities requires specific prerequisites and could allow a sophisticated attacker to compromise the security of the communication devices, allowing an attacker to escalate privileges, view sensitive information, upload arbitrary files, and perform remote code execution.
Under very specific conditions, successful exploitation of these vulnerabilities could allow an attacker to modify the configuration of a connected infusion pump Perfusor®, Infusomat®, or Infusomat® P from the Space and compactplus product family while the pump is not delivering therapy.
These conditions include the following: (i) the attacker has already gained access to this specific hospital network, (ii) the affected pumps are connected to a hospital network, (iii) the attacker specifically targets the affected device using the identified vulnerabilities, (iv) the infusion pump is not delivering a therapy (it is “Turned Off” or in “Standby Mode”).
These vulnerabilities do not allow an attacker to modify a therapy while the infusion pump is actively delivering therapy!
B. Braun has received no reports of exploitation or incidents associated with these vulnerabilities in real-world use environment.
The following supported B. Braun products are affected:
Other components of B. Braun Space, compactplus and Spaceplus infusion pump systems are not affected. Installations of Space or compactplus infusion pumps without the above listed communication devices are not affected.
B. Braun SpaceCom and Datamodule compactplus use the 3rd-party package ‘PHP’. This 3rd-party package has the known vulnerabilities CVE-2026-109618 and CVE-2026-109619. The issue allows an attacker to infiltrate the product via the network.
CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
COUNTRIES/AREAS DEPLOYED: SpaceCom: worldwide; compactplus: selected countries
COMPANY HEADQUARTERS LOCATION: Germany
Internal product surveillance discovered this 3rd-party vulnerability to affect B. Braun solutions.
The following versions include a patch for this vulnerability:
As a general security measure, B. Braun recommends applying established network security practices appropriate to the healthcare facility’s network architecture and risk assessment. These include network segmentation and access controls that limit device access to authorized systems and users. The affected communication devices should not be directly accessible from the internet or untrusted networks.
Wireless networks should use appropriate industry-standard encryption and authentication. Additional controls such as firewalls, VLANs, network monitoring, and intrusion detection or prevention may be applied where appropriate.
These measures reduce network exposure and complement, but do not replace applicable software updates and product-specific remediation measures.
Please contact our global security team at productsecurity@bbraun.com if you have any further questions, require detailed technical information, or have any other support issue concerning the security of our products and services.
If you are a B. Braun customer and need support, please contact your local B. Braun representative.
This is the global B. Braun vulnerability disclosure. Customers in the United States and customers in Canada using the software ‘U’ should refer to the U.S.-specific vulnerability disclosure available at www.bbraunusa.com/productsecurity.
Your feedback matters! Participate in our customer survey to help us enhance our website, products and services. Thank you for your support!